Mold incubator use method

Fill the water tank with pure water according to the instructions, insert the ultrasonic atomizer power cord into the rear panel diagram (2) " humidifier control socket", insert the soft plastic sleeve at the exit of the ultrasonic atomizer, and One end is inserted in the top plate schematic (3) "air intake", at this time the ultrasonic atomizer has been connected to the host.

5. Plug in the power of the host and press the power switch “1” on the right side of the panel. At this time, the instrument displays a digital display, indicating that the device enters the working state. See the control panel for each parameter setting. The setting method is as follows:

(1) Temperature/time switch for temperature and time switching:

1) Temperature setting: Press the temperature/time button and the temperature setting indicator of the meter will light. Press the SET button to enter the SV state, then press the ∨ or ∧ button to change the temperature set value to the desired temperature (minimum 0 °C). Finally, press the SET button to save the set value to the MCU. The window displays the measured temperature and the right display shows the set temperature.

2) Time setting: Press the temperature/time button, the meter's time setting indicator lights. Press the SET button to enter the TV state and then press the ∨ or ∧ button to change the time setting value until the required time (minimum 1h), and finally press the SET button to save the set value to the MCU, then the left display window The measurement time is displayed, and the right display window shows the set time.

Timing function: When the instrument is powered, the timing function starts. After the timing is over, the output is turned off. The heater and the refrigeration compressor stop working, and the temperature inside the incubator will gradually reach the ambient temperature.

3) Status indication: After the setting is completed, the device runs according to the set parameters. When the measured temperature is lower than the set temperature, the device enters the heating state and the heating indicator lights. Conversely, when the measured temperature is higher than the set temperature, the cooling indicator lights up and the device enters the cooling state. When the measured temperature is higher than the set temperature by 4 °C, the alarm indicator lights and the buzzer sounds.

4) Parameter modification: If you want to modify the temperature and time parameters during operation, modify it according to the procedures 1 and 2.

5) Quick setting: When setting parameters, one of the three numbers will flash. If you press the ∧ key, the last digit of the set value will increase by one; if you press this button, it will not open, after about 3 seconds, set The set value is continuously increased by 1; if the ∨ key is pressed, the set value is decreased by 1; if the hold is not released, the set value is continuously decreased by 1 after about 3 seconds.

(2) Humidity setting

1) Humidity setting: press the “humidity setting” button, the “upper limit setting” indicator (red) of the window will be on. Press the ∨ or ∧ key to change the humidity setting value to the required humidity, and then press Under the “humidity setting” button, the “lower limit setting” indicator (green) of the window is on. Press the ∨ or ∧ key to set the lower limit humidity, then press the “humidity setting” button to “work” the window. The indicator light (red) is on, indicating that the microcomputer is in automatic operation. The "lower limit setting" humidity value differs from the "upper limit setting" humidity value by 2% (recommended value 1% to 5%), and the user can also set it according to the actual situation.

2) Parameter modification: If you want to modify the humidity parameter during operation, modify it according to program 1.

3) Quick setting: When setting the parameters, if you press the ∧ key, the last digit of the set value will be increased by 1. If you press this button, the setting value will be incremented by 1 after about 3 seconds; if you press ∨ If the key is pressed, the set value is decremented by 1. If the key is pressed and not released, the set value is continuously decreased by 1 after about 3 seconds.

4) status indication: When the humidity inside the humidity does not reach the lower limit is set, the indicator light and humidification, humidification nebulizers work, humidity rises to set the lower limit of the humidity after the humidifying intermittently, until the stop limit humidity Humidification. Allow the humidity to fluctuate within the required range. When the humidity meter displays EFF, it is an overflow display, and the actual humidity is outside the hygrometer display range. At this point, the outer door of the incubator can be opened, and the display will automatically resume after a period of time. If the humidity is low, it can be humidified.